95.2 and 98.8 are 3.6 percentage points apart, which sounds small until you restate it as impurity. PeptideDetective is leaving 4.8 per cent of the detected area unassigned and Janoshik 1.2 per cent — a factor of 4 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 95.2 and 98.8 are two measurements of slightly different quantities that happen to share a unit.
Mechanically, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
The part that matters: integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.